3mm Cannulas and Trocars

Home » 3mm Cannulas and Trocars

YelloPort+plus™ 3mm - 'Ultra' Minimally Invasive Surgery

3mm cta image

The aim of 3mm surgery is to substantially reduce post-operative pain and trauma for the patient, promote faster recovery time and improve ‘cosmesis’ (minimal scarring) by making smaller incisions. Contact us today!

Key Strengths of YelloPort+plus™ 3mm

  • Impressive strength despite reduced size
  • Lightweight and small profile in construction
  • Cannula cap designed to allow easy instrument insertion
  • Perfect complement to Logi™Range 3mm Resposable® instruments

YelloPort+plus™ 3mm Obturator Options

Product Name

Product Code

Working Length

Pyramidal Obturator

pyramind

YT0305501

YT0307501

YT0309501

55mm

75mm

95mm

Pencil Point Obturator

pencil

YT0305503

YT0307503

YT0309503

55mm

75mm

95mm

YelloPort+plus™ 3mm Cannulae Options without Thread

Product Name

Product Code

Working Length

Cannula

3mmnothreadNoluer

YC0305500

YC0307500

YC0309500

55mm

75mm

95mm

Cannula + Luer

3mmNoThrdLuer

YC0305501

YC0307501

YC0309501

55mm

75mm

95mm

YelloPort+plus™ 3mm Cannulae Options with Thread

Product Name

Product Code

Working Length

Cannula + Thread

3mmthreaded

YC0305510

YC0307510

YC0309510

55mm

75mm

95mm

Cannula + Luer + Thread

3mmthreadedLuer

YC0305511

YC0307511

YC0309511

55mm

75mm

95mm

YelloPort+plus™ 3mm Consumables




Product Description

Sterile valve

3mmValve

YA03VSS01

1 per pack

Luerlock Cap

5mmluerlockcap

YALLC10

10 per pack


More Information About Minimally Invasive Surgery

Minimally invasive surgery (MIS), commonly known as 'keyhole' surgery, involves entering the skin through small incisions with the aim of ensuring:

  • Less post-operative pain and trauma
  • Faster recovery
  • Improved cosmesis (cosmetic appearance of scarring)
  • Less invasive than open surgery
  • DrAmado Marco Adamo, Consultant Surgeon
  • “I have found YelloPort+plus™ 3mm ports extremely good, especially for use in laparoscopic hernia repairs. In my opinion, they are as good as the conventional 5mm instruments, just smaller. The fact that they are smaller means they require smaller incisions and the surgical dissection is more precise.”


Cleaning and Sterilization:

YelloPort Trocars and Cannulae are suitable for cleaning in Commercial Washing / Disinfecting machines.

All components capable of sterilizaton in pressurised steam using standard autoclave cycles.

Flash Sterilizaton Not Recommended. It shortens product life.

With Dynamic Air-Removal steam sterilization: Porous Load. Yelloports are "Wrapped Instruments".
Exposure Temperature = 132 - 137C with dynamic air-removal steam sterilization: porous load. Wrapped Instruments.

Exposure Time = 4 minutes.


With Gravity Displacement Steam Sterilization, Non-Wrapped Instruments. Yelloports are "Unwrapped Non-Porous Items".

Recommended Sterilizaton Temperature = 132 - 137C.
Exposure Time = 15 minutes.

DO NOT sterilize in Ethylene Oxide (EO or ETO).

La Salle International Ltd 800-580-6747

Email: YelloPort@yahoo.com